Forsker
- Bergen
stst@norceresearch.no
+47 56 10 78 24
NORCE Multiphase Flow Loop
A full scale oil-water-gas flow loop
The NORCE Multiphase Flow loop is a full scale oil-water-gas flow loop owned and operated by NORCE, located at Fantoft in Bergen.
The facility was designed and built by Christian Michelsen Research (CMR), which merged into NORCE in 2018.
Collaborate with us
NORCE as well as external companies use the flow loop for R&D and third party tests of flow instruments.
The test facility consists of a 3” and a 6” flow line, a 32 m3 separator tank, centrifugal pumps and reference instrumentation for determination of flow rates, phase fractions (WLR, GVF, etc.), fluid densities, flow pressures and temperatures.
The reference instruments are subject to traceable, accredited calibration according to ISO 17025. The fluids used are diesel, nitrogen and salted (NaCl) tap water.
The facility has been used for more than 200 factory acceptance tests (FATs) of subsea and topside multiphase flow meters (MPFM) since start-up in 2009.
Contact us if you want to work wit us.
Specifications
- Separator pressure: 5 bara
- Flow pressure: 5 - 8 bara, depending on flow rate
- Temperature range: 15 - 35 °C (no temperature regulation)
- Gas flow rate: 1100 Sm3/h
- Liquid flow rate: 100 m3/h in 3’’ loop and
250 m3/h in 6’’ loop - Water liquid ratio: 0 - 100 %
- Salinity: 1 - 8 wt.% NaCl (20-120 mS/cm)
Documentation of operational procedures and the uncertainty of reference instrumentation is available upon request.
Visiting address
Fantoftvegen 38, Bergen, Norway
Post
NORCE Flerfaserigg
POB 22 Nygårdstangen, NO-5838,
Bergen Norway